Problem |
Solution |
Error Code 1 (power interruption
to sterilizer)
1. Unit powered up for first time.
2. Power was interrupted by power outage.
3. Overheat switch was activated.
|
1. Press ON/STANDBY button and then
continue normal operation.
2. Press ON/STANDBY button and then continue normal
operation.
3. Allow sterilizer to cool completely. Restart
sterilizer. If unit has errors again, unplug unit and
contact Alfa.
|
Error Code 2 (STOP button pushed
during cycle) |
Restart sterilizer after unit stops
beeping. |
Error Code 3 (ON/STANDBY button
pushed during cycle) |
Restart sterilizer after unit stops
beeping. |
Error Code 4 (door ajar during
cycle)
1. Door was opened during cycle.
2. Door switch not functioning properly.
|
1. Restart sterilizer after unit
stops beeping.
2. Allow sterilizer to cool completely. Restart
sterilizer. If unit has errors again, unplug unit and
contact Alfa.
|
Error Code 5 (excessive pressure)
1. Sterilizer overloaded with gauze or linen packs.
2. Unit is out of calibration or bellows is not
functioning properly.
|
1. Reload the sterilizer in
accordance with Guidelines for Loading Trays. If the problem
recurs, take the unit out of service and contact Alfa.
2. Allow sterilizer to cool completely. Restart
sterilizer. If unit has errors again, unplug unit and
contact Alfa.
|
Error Code 6 (excessive
temperature)
1. Unit is out of calibration.
|
1. Allow sterilizer to cool
completely. Restart sterilizer. If unit has errors again,
unplug unit and contact Alfa. |
Error Code 7 (pressure low during
cycle)
1. Leaking door gasket.
2. Internal leak in unit plumbing.
|
1. Allow sterilizer to cool; then
clean or replace door gasket (check with your manual.
2.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|